FDA Device Recalls

Recalls /

#122995

Product

Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.

FDA product code
JQPCalculator/Data Processing Module, For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
Affected lot / code info
CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474.

Why it was recalled

Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)

Root cause (FDA determination)

Software design

Action the firm took

The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated October 23, 2013 to its customers. The firm also sent a follow-up letter dated October 30, 2013. The letter describes the product, problem and actions to be taken. The customers were instructed Until a service visit is scheduled, do the following: Instruct the laboratory staff to recognize this sequence of events and Ensure that whenever the operator disables a method on an Aptio system or a CentraLinksystem and replenishes reagent for that method, the operator then confirms that the method remains disabled in the CentraLink system or disables it again. The operator must allow up to five minutes to verify that the test remains disabled. In addition, the customers were instructed tocomplete and return the FIELD CORRECTION EFFECTIVENESS CHECK response form via fax to Siemens Healthcare Diagnostics Customer Care Center at 800-441-9250, and to retain this letter with your laboratory records, and forward this letter to those who may have received this product. After a service visit is performed and an interim solution is implemented, the laboratory staff will be trained on this update to understand the minor but necessary changes in the workflow. If you have any questions, please contact your Siemens Customer Care Center at 800-441-9250 or your local Siemens representative.

Recalling firm

Firm
Siemens Healthcare Diagnostics, Inc.
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466

Distribution

Distribution pattern
US Distribution in the states of: MN, NJ, and NV.

Timeline

Recall initiated
2013-10-30
Posted by FDA
2014-01-24
Terminated
2014-12-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #122995. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.