Product
Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain Dual patient lines; Product Code: 3652-100. To evacuate air and/or fluid from the chest cavity or mediastinum.
- FDA product code
- CAC — Apparatus, Autotransfusion
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K043140, K043582
- Affected lot / code info
- Lot number starting with 10883097 and the ending lot number 10913346. All units in between and including these two numbers.
Why it was recalled
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
Root cause (FDA determination)
Device Design
Action the firm took
Atrium Medical issued notification letters to direct consignees by UPS next day on 19 November 2013, and indirect consignees (distributors customers) by UPS next day or USPS priority (for customers with a PO Box) on 27 November 2013. The Letter explain the issue and the steps that should be taken. The customer will be offered replacement with unaffected product or credit. The firm describes Procedures for Emergency Use. The firm identifies potential product substitutions. These products are the same drainage product without the auto-transfusion line. International Distributors. International distributors provided with the Recall Authorization' package and the customer letter. Accounts are requested to complete the Recall Reply Form acknowledging receipt of the notification. Contact Atrium Medical Customer Service at 1-800-370-7899, Monday through Friday between 9:00 am to 5:00 pm.
Recalling firm
- Firm
- Atrium Medical Corporation
- Address
- 5 Wentworth Dr, Hudson, New Hampshire 03051-4929
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.
Timeline
- Recall initiated
- 2013-11-19
- Posted by FDA
- 2014-01-14
- Terminated
- 2015-06-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #123497. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.