Recalls / —
—#123691
Product
4040-2: Umbilical 3 mL Line Draw, Luer lock syringe (heparin concentration: 23.5 I.U. per mL) with Filter-Pro device, 22g x 1 needle, and Point-Lok needle safety device Arterial Blood Sampling Line Draw Syringes are sterile, single-use devices, in-vitro diagnostic devices used for the drawing of arterial blood via an arterial line. The product is not for injection.
- FDA product code
- JKA — Tubes, Vials, Systems, Serum Separators, Blood Collection
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K952516
- Affected lot / code info
- 2485103 2493959 2498064 2511207 2521581 2525780 2525781 2533316 2547272 2574207 2578556 2472508 2476319 2467728 2497730
Why it was recalled
issue with the plunger tip of the Portex Arterial Blood Sampling Line Draw Syringes
Root cause (FDA determination)
Component change control
Action the firm took
Smiths Medical sent an Urgent Field Safety Notice dated November 22, 2013, to all affected consignees via FedEx beginning on November 23, 2013. Distributors were instructed to notify their customers. Customers were instructed to inspect their inventory for the affected product and quarantine the affected syringes. Complete and return the attached Urgent Field Safety Notice Confirmation Form by fax to 603-358-1017 or by email to LineDraw@smiths-medical.com. Upon receipt of the completed form, a customer service representative will contact them to arrange for exchange of their unused affecte devices for credit or replacement. Customers were instructed to forward the notice to all personnel who need to be aware within their organization and to any organization where the potentially affected devices have been transferred. Customers were instructed to report any issues with the product to Smiths Medical's Global Complaint Department at 1-866-216-8806 or globalcomplaints@smiths-medical.com. For questions regarding this recall call 603-352-3812.
Recalling firm
- Firm
- Smiths Medical ASD, Inc.
- Address
- 10 Bowman Dr, Keene, New Hampshire 03431-5043
Distribution
- Distribution pattern
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV. Internationally to Australia, Austria, Belgium, Belize, Brazil, Canada, Czech Republic, Denmark, England, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Netherlands, Norway, Poland, Scotland, Spain, and Sweden.
Timeline
- Recall initiated
- 2013-11-22
- Posted by FDA
- 2014-03-12
- Terminated
- 2020-11-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #123691. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.