Product

Dimension Vista Cyclosporine Flex reagent cartridge an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista System. Measurements are used as an aid in the management of heart, liver, and kidney transplant patients.

FDA product code
MKWCyclosporine
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K062236
Affected lot / code info
lots 12300BB, exp. 10-26-2013; 12318BB, exp. 11-13-2013; and 13011BB, exp. 1-11-2014.

Why it was recalled

Complaints were received regarding variability in recovery of QC and patient results with certain lots of the Dimension Vista CSA Flex reagent cartridges.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Siegmens sent an Urgent Medical Device Recall lletter dated December 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use and discard any remaining inventory of the affected product. Customers were also instructed to complete the attached form and fax it to 302-631-8467. Customers were asked to retain the letter with their laboratory records, and forward the letter to any who have received the affected product. Customers with questions were instructed to contact Siemens Customer Care Center - Technical Solutions or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.

Recalling firm

Firm
Siemens Healthcare Diagnostics, Inc.
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Canada.

Timeline

Recall initiated
2013-12-05
Posted by FDA
2014-01-23
Terminated
2014-08-27