Recalls / —
—#125055
Product
Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.
- FDA product code
- JIX — Calibrator, Multi-Analyte Mixture
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K061923
- Affected lot / code info
- lot 3FD034, exp. 6/17/2014, and lot 3HD012, exp. 9/1/2014.
Why it was recalled
Complaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) and imprecision during PCHE calibration with the Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL).
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
A recall notification letter, dated January 2014, was sent to Users.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- Nationwide Distribution.
Timeline
- Recall initiated
- 2014-01-06
- Posted by FDA
- 2014-02-24
- Terminated
- 2015-05-06
- Status
- —