Product

Sutures, Absorbable, Synthetic, Polyglycolic Acid, Sterile, Rx only, Product Usage: Non-absorbable polypropylene surgical suture is a monofilament, Non-absorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation.

FDA product code
GATSuture, Nonabsorbable, Synthetic, Polyethylene
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K021019
Affected lot / code info
Product Code: BON100, Lot numbers: 02H1302839, 02J1101705, 02D1101137, and 02F1103013.

Why it was recalled

The products are being recalled because they did not meet minimum needle attachment strength requirements.

Root cause (FDA determination)

Employee error

Action the firm took

Teleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990

Recalling firm

Firm
Teleflex Medical
Address
2917 Weck Dr., Research Triangle Park, North Carolina 27709

Distribution

Distribution pattern
Worldwide Distribution - U.S. Nationwide and the countries of Canada, Germany, and Ireland

Timeline

Recall initiated
2014-03-11
Posted by FDA
2014-03-26
Terminated
2017-08-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #126183. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.