FDA Device Recalls

Recalls /

#127246

Product

VITROS 5600 Integrated System, Product Code/Catalog Number 6802413 for in vitro diagnostic use.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K081543
Affected lot / code info
Software Version 3.0 and lower; Serial Numbers J56000110 through J56001912

Why it was recalled

Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with results from any of the following VITROS Chemistry Products using multiple point (multipoint) rate measurements. The MicroSlide assays that could be affected

Root cause (FDA determination)

Software design

Action the firm took

Important Product Correction Letters (CL2014-116, dated 4/10/2014) were sent on 4/10/2014 via FedEx overnight courier or ORTHO PLUS e-Communications to all customers with VITROS 5,1 FS, 5,1 FS Refurbished, 4600,and/or 5600 Systems. They were informed of the issue and advised to: -- Review VITROS ALT, ALTJ, ALKP, ALKPJ, AST, ASTJ, CHE, CHEJ, CK, CKJ, GGT, GGTJ, LDH and LDHJ results generated on VITROS 4600/5600 and 5,1 FS Systems. Do not report any results associated with an MW code. -- Perform repeat testing in singlicate on all samples with results having an MW code. Only results without an MW code are acceptable. -- Review their current Laboratory Information System (LIS) configuration to determine if changes are required in order to restrict reporting of results with an MW code. Customers with questions should contact Customer Technical Services at 1-800-421-3311. --- Foreign affiliates were informed of the issue by e-mail on 4/10/2014 and instructed to notify their consignees of the issue and actions.

Recalling firm

Firm
Ortho-Clinical Diagnostics
Address
100 Indigo Creek Dr, Rochester, New York 14626-5101

Distribution

Distribution pattern
Worldwide distribution including US Nationwide, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

Timeline

Recall initiated
2014-04-10
Posted by FDA
2014-06-09
Terminated
2018-05-21
Status

Source: openFDA Device Recall endpoint. Recall record ID #127246. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: VITROS 5600 Integrated System, Product Code/Catalog Number 6802413 for in vitro diagnostic use. · FDA Device Recalls