Product

NexGen Complete Knee Solution, MIS Total Knee Procedure, Stemmed Tibial Component, Precoat, Fixed Bearing, Size 1. Part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis.

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K043101
Affected lot / code info
Size 1, includes the following Lot numbers: 37217848 61424993 61507222 61543969 61774117 62141262 62193298 62251564 62261955 62383353 62602809 62602810 62602811

Why it was recalled

Zimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.

Root cause (FDA determination)

Equipment maintenance

Action the firm took

Zimmer sent an Urgent Medical Device Recall letter June 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification and ensure affected personnel are aware of the contents. Locate and quarantine any affected product and notify their Zimmer sales representative. Their Zimmer sales representative will remove the recalled product from their facility. For patients that previously had this product implanted, it is recommended that they continue their normal operative follow up routine. Customers with questions were instructed to call 1-877-946-2761.

Recalling firm

Firm
Zimmer, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Australia, China, Hong Kong, India, Japan, Korea, Malaysia, Nicaragua, Singapore, and Taiwan.

Timeline

Recall initiated
2014-05-19
Posted by FDA
2014-06-27
Terminated
2015-10-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #127775. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.