Product
ADVIA Centaur Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), SMN 10317708 and 02790309 (500 tests), SMN 10317790
- FDA product code
- MMI — Immunoassay Method, Troponin Subunit
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K053020
- Affected lot / code info
- Kit Lots Ending In: 078, 079, 082, 083, 084
Why it was recalled
The solid phase reagent in some of the ADVIA Centaur TnI -Ultra ReadyPacks, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
An Urgent medical Device Correction (UMDC) was mailed, via FedEx, to all affected Siemens Healthcare Diagnostics customers in the United States on May 30, 2014 for customer receipt on June 2, 2014. All international customers were provided with a copy of the Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each country's local regulations and procedures. Each customer is asked to complete a fax-back form. This notice informed the customers of the potential for incorrect control and patient results due to solid phase which is darker in appearance. The notice instructs customers to place one TnI-Ultra ReadyPack on the system at a time and to calibrate and run controls on each ReadyPack to ensure accurate control and patient values. Patient TnI-Ultra results produced with this assay are acceptable if they follow acceptable calibration with valid quality control results. Upon receipt of kit lots ending in 086 and higher, calibration of each ReadyPack is not required. This notice should be reviewed with the facility's laboratory or medical director.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc
- Address
- 333 Coney St, East Walpole, Massachusetts 02032-1516
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) including PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada, Mexico, Argentina, Austria, Australia. Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Great Britain, Greece, Guadeloupe, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Libya, Macedonia, Malaysia, ,Malta, Martinique, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Republic of Korea, Romania, Russian Federation, Serbia, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sri Lanka, South Africa, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Vietnam.
Timeline
- Recall initiated
- 2014-05-30
- Posted by FDA
- 2014-07-02
- Terminated
- 2016-02-26
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #128130. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.