FDA Device Recalls

Recalls /

#129074

Product

Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE

FDA product code
HRSPlate, Fixation, Bone
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K830196
Affected lot / code info
Item Numbers: 00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514      EXPIRATION DATE PRIOR TO 2024-07

Why it was recalled

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Root cause (FDA determination)

Package design/selection

Action the firm took

Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.

Recalling firm

Firm
Zimmer, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Worldwide Distribution.

Timeline

Recall initiated
2014-07-22
Posted by FDA
2014-08-26
Terminated
2015-09-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #129074. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.