Recalls / —
—#129137
Product
Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
- FDA product code
- JQP — Calculator/Data Processing Module, For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- Affected lot / code info
- EasyLink Informatics System V4.0/V5 Siemens Material Number (SMN) 10444799, 10444800, 10487219
Why it was recalled
May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Root cause (FDA determination)
Software design
Action the firm took
Siemens issued an Urgent Medical Device Correction (UMDC) letter on August 13, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review this letter in its entirety and ensure that all operators understand the information presented. The issues described are specific to the features listed in the letter.. If the laboratory is not using a particular feature, then they are not affected by the issues described for that feature. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- US Nationwide Distribution
Timeline
- Recall initiated
- 2013-08-13
- Posted by FDA
- 2014-09-01
- Terminated
- 2014-12-02
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #129137. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.