Recalls / —
—#129447
Product
Gown Cool Impervious XLGE Breathable Customed, 900-2592: (1) GOWN XL SURGICAL MICROCOOL (1) WRAPPER 24" X 24" (1) TOWELS ABSORBENT 15" X 20" LIF Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- FDA product code
- LRO — General Surgery Tray
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- Product code 900-2592, 20 lot: 112020341 112041064 112041312 112051974 112062485 112062496 112124666 112125076 113057401 113067920 113078313 113088800 113089085 113099304 113109818 131110252 131210794 140111212 140312135 140412659
Why it was recalled
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root cause (FDA determination)
Package design/selection
Action the firm took
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Recalling firm
- Firm
- Customed, Inc Calle Igualdad Final # 7
- Address
- Fajardo, Puerto Rico 00738
Distribution
- Distribution pattern
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Timeline
- Recall initiated
- 2014-05-20
- Posted by FDA
- 2014-08-29
- Terminated
- 2017-08-14
- Status
- —