FDA Device Recalls

Recalls /

#129540

Product

ORTHOPEDIC PACK BASIC CODE 900-2830. CONTENTS: (3) TOWELS ABSORBENT 15" X 20" (1) COVER MAYO STAND REINFORCED (1) U-DRAPE 60" X 70" WITH TAPE SPLIT 6 X 21 SPLIT (1) DRAPE% ECONOMY 53" X 77" (4) DRAPE UTILITY WITH TAPE (1) BAG SUTURE FLORAL (1) SHEET SPLITW/ADHES 108" X 77" STD SMS (1) COVER TABLE REINFORCED 50" X 90" (1) STOCKINETTE IMPERVIOUS 14" X 48" (1) TIME OUT BEACON NON WOVEN (1) GOWN XL SMS IMPERVIOUS REINFORCED AAMI LEVEL Ill (1) POLISHER CAUTERY TIP (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER (1) COUNTER NDLIBLADE 20C FOAM/MAG (1) PENCIL CAUTERY ROCKER SWITCH WITH HOLSTER (1) GOWN IMPERVIOUS EXTRA REINFORCED LGE T/WRAP AAMI LEVEL Il (1) BAG GLASSINE PLAIN (1) ELECTRODE DUAL DEPRES Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

FDA product code
OJHOrthopedic Tray
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Code 900-2830, 3 lots 140111332 140211832 140412672

Why it was recalled

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Root cause (FDA determination)

Package design/selection

Action the firm took

Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540

Recalling firm

Firm
Customed, Inc Calle Igualdad Final # 7
Address
Fajardo, Puerto Rico 00738

Distribution

Distribution pattern
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Timeline

Recall initiated
2014-05-20
Posted by FDA
2014-08-29
Terminated
2017-08-14