Recalls / —
—#129543
Product
BASIC COSMETIC SURG. PACK CUSTOMED (USA), CODE 900-2915. CONTENTS: (2) TOWELS ABSORBENT 15" X 20" LIF (1) DRAPE BREAST/CHEST SMS L/F (1) SYRINGE 20cc W/0 NEEDLE LUER LOCK LIF (1) TUBE SUCTION CONNECT Yt" X 12' LIF (1) YANKAUER SUCTION TUBE W/0 VENT L/F (1) STERI STRIP CLOSURE%" X 4" L/F (1) BLADE SURG. #15 CARBON STEEL (1) NEEDLE HYPODERMIC 25 X 1 % LIF (5) LAP SPONGE PREWASH 18" X 18" XRD LIF (1) PENCIL CAUTERY ROCKER SWITCH W/ HOLSTER LIF (1) SYRINGE 60cc THUMB CONT. RING CATH. TIP LIF (1) COVER TABLE REINFORCED 50" X 90" LIF (1) COVER MAYO STAND REINFORCED LIF (1) BAG SUTURE FLORAL (2) GOWN SOFT SMS STD. XL SET IN SLEEVE LIF (4) DRAPE UTILITY WITH TAPE L/F (4) TOWEL CLOTH HUCK LIF (1) WRAPPER 24" X 24" Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- FDA product code
- LRO — General Surgery Tray
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- Code 900-2915, 4 lots 113026307 113047021 113068035 113078731
Why it was recalled
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root cause (FDA determination)
Package design/selection
Action the firm took
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Recalling firm
- Firm
- Customed, Inc Calle Igualdad Final # 7
- Address
- Fajardo, Puerto Rico 00738
Distribution
- Distribution pattern
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Timeline
- Recall initiated
- 2014-05-20
- Posted by FDA
- 2014-08-29
- Terminated
- 2017-08-14
- Status
- —