Recalls / —
—#129993
Product
VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
- FDA product code
- JJE — Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K081543
- Affected lot / code info
- Software Version 3.0 and below; Serial Numbers J56000110 - J56001995 (serial numbers are sequential; J Numbers are analogous to serial numbers)
Why it was recalled
Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).
Root cause (FDA determination)
Software design
Action the firm took
On 9/08/14, URGENT PRODUCT CORRECTION NOTIFICATION Letter (CL2014-232, dated 9/8/14) and Confirmation of Receipt Form were sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government only) to inform them of the issue and advise them of the required actions. Foreign affiliates were informed of the issue via e-mail on 9/8/14 and instructed to notify their consignees of the issue and the required actions.
Recalling firm
- Firm
- Ortho-Clinical Diagnostics
- Address
- 100 Indigo Creek Dr, Rochester, New York 14626-5101
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Timeline
- Recall initiated
- 2014-09-08
- Posted by FDA
- 2014-10-16
- Terminated
- 2018-05-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #129993. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.