FDA Device Recalls

Recalls /

#130144

Product

ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyethylene bag

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
Catalog NOs. 2544-00-521 and 2544-00-520; All lots are affected for the Device Correction.   Lot numbers below should be inspected and swapped out if the bushing is incorrectly assembled: Part No. DePuy Synthes Lot No: MFG Lot No: 254400520 ABC22844 C22844  ABC22845 C22845  ABC34279 C34279  ABC34280 C34280  ABC4107 C4107  ABC4108 C4108  ABC4112 C4112  ABC4113 C4113  ABC50741 C50471 254400521 ABB13662 B13662  ABB19116 B19116  ABB19117 B19117  ABB19118 B19118  ABB35392 B35392  ABB35393 B35393  ABB39835 B39835  ABB41384 B41384  ABB46872 B46872  ABB49200 B49200  ABB50724 B50724  ABB55749 B55749  ABB63598 B63598  ABB66346 B66346  ABB72107 B72107  ABB77183 B77183  ABB79390 B79390  ABB79391 B79391  ABB80712 N80712  ABC17165 C17165  ABC5023 C5023

Why it was recalled

Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.

Root cause (FDA determination)

Process control

Action the firm took

DePuy sent an "URGENT DEVICE CORRECTION" notifications dated October 15, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, please call 574-371-4917.

Recalling firm

Firm
DePuy Orthopaedics, Inc.
Address
700 Orthopaedic Dr, Warsaw, Indiana 46582-3994

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) except AK, including Puerto Rico and the countries of Canada, Australia, Austria, Belgium, China, France, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, UAE, United Kingdom and Chile.

Timeline

Recall initiated
2014-10-15
Posted by FDA
2014-11-10
Terminated
2015-11-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #130144. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.