FDA Device Recalls

Recalls /

#130644

Product

Wound Closure Tray I, Catalog number 900-063. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

FDA product code
MCYWound Dressing Kit
Device class
Class 2
Medical specialty
General Hospital
Affected lot / code info
Lot numbers: 111112932 111123285 112010264 112030760 112041053 112051645 112062307 112083045 112093740 112104126 112114893 113025998 113046869 113078228 113078662 113099263 131110063 131210669 140111107 140211528 140412569 140412798 140613556 140714140 140815029

Why it was recalled

Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.

Root cause (FDA determination)

Packaging

Action the firm took

Customed initiated a recall on October 8th, 2014 of sterile convenience surgical Packs-trays, bags (known as Convenience Packs) and new lots of products added to the initial June 3, 2014 recall of 233 sterile surgical packs. The products are being recalled because of significant compromised sterility, not limited to potentially damaged packaging. Customers were instructed to quarantine and cease use of the affected product identified. Customers are to return the response form included with the letter, to the contact information provided. Upon receipt, the firm will contact customers with further instructions to obtain credit, replacement product, and for return of the recalled product or any other disposition action. If customers have any questions, they should contact the Customed Recall Coordinator at 407-850-5558 extensions 8001 or 8005, or via e-mail: pablo.diaz@customedhealing.com.

Recalling firm

Firm
Customed, Inc Calle Igualdad Final # 7
Address
Fajardo, Puerto Rico 00738

Distribution

Distribution pattern
US Distribution to Florida, New York and Puerto Rico..

Timeline

Recall initiated
2014-10-08
Posted by FDA
2014-12-16
Terminated
2017-08-14
Status