FDA Device Recalls

Recalls /

#131032

Product

Siemens SimView NT and SimView NT with IDI simulation system components: Product Usage: The intended use of the Radiation Therapy Simulation system is used for simulation of treatment planning prior to the therapeutic treatment of cancer.

FDA product code
KPQSystem, Simulation, Radiation Therapy
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K953233
Affected lot / code info
Model numbers: 5496992, 5497016   with Serial numbers: 6035, 6080, 6029, 6032, 6019, 6001

Why it was recalled

Siemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that simulation system components could collide with patient and/or system equipment.

Root cause (FDA determination)

Device Design

Action the firm took

Customers were initially notified about the issue by letter on January 7, 2014 (this action was reported as Z-0786-2014 which has been terminated). Siemens promised to notify them when a fix became available. Customer letter, dated November 7, 2014, was sent to notify end users that a fix has become available and Siemens would contact them to schedule an appointment for the correction.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355

Distribution

Distribution pattern
Worldwide Distribution - US Nationwide in the states of PA, KY, LA, MA, MI, and CA; and countries of: Argentina, Belarus, Bulgaria, Costa Rico, P.R. China, Curacao, Egypt, Repulic Korea, Germany, Greece, Hungary, Italy, Iraq, Jordan, Malaysia, Palestanian Ter, Poland, Thailand, Turkey, and Vietnam.

Timeline

Recall initiated
2014-10-10
Posted by FDA
2014-11-20
Terminated
2017-06-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #131032. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.