Recalls / —
—#131283
Product
Aquapak Adaptor, 028 NEB, INTL, Product No. 403128; Aquapak Adaptor, 033 NEB, INTL, Product No. 403133; Aqaupak Nebulizer Adaptor 028, Sterile, Shelfpak, Product No. 031-28 & 031-28F, Aqaupak Nebulizer Adaptor 033, Sterile, Shelfpak, Product No. 031-33; Aqaupak Nebulizer Adaptor 033, Sterile, Japanese, Product No. 031-33J. Used in anesthesiology.
- FDA product code
- CAF — Nebulizer (Direct Patient Interface)
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- Affected lot / code info
- Product No. 043128, Lot # 02E1200719, 02E1200720, 02E1201281, 02F1200964, 02K1201946, 02E1201808, 02E1203097, 02G1300256, 02G1301048, & 02M1201691; Product No. 403133, Lot # 02E1201282, 02K1201948, 02F1200965, 02L1201194, 02M1200230, & 02F1302063; Product No. 031-28, Lot # 02D1202234, 02A1202383, 02A1202382, 02B1201069, 02D1200247, 02D1200248, 02D1201944, 02D1202274, 02D1202733, 02D1202236, 02D1202734, 02D1202735, 02D1202233, 02G1200664, 02G1200678, 02G1201016, 02G1201698, 02F1200961, 02G1200676, 02M1200585, 02L1200927, 02M1200749, 02M1201693, 02A1300194, 02A1300457, 02D1301655, 02D1301191, 02D1300864, 02D1301192, 02D1302730, 02F1301742, 02E1302765, 02F1300739, 02F1302062, & 02D1200249; Product No. 031-28F, Lot # 02E1200717, 02K1201947, 02F1200962, & 02E1202197; Product No. 031-33, Lot # 02E1200718, 02A1202384, 02E1202198, 02F1200963, 02G1200666, 02J1202240, 02M1200584, 02M1200748, 02M1201771, 02D1302731, 02A1300458, 02B1300718, & 02M1201692; Product No. 031-33J, Lot # 02D1202258, 02D1202260, 02D1202259, 02D1202261, 02E1201563, 02E1201279, 02E1203866, 02M1200747, 02B1300717, 02B1301203, 02E1301554, 02B120342, 02J1201473, 02J1202736, 02F1201179, & 02M1200057.
Why it was recalled
Possibility that the nebulizer adaptor packaging may have open or weak seals which may affect the sterility of the adaptor.
Root cause (FDA determination)
Packaging change control
Action the firm took
Customers were notified of the recall by letter on September 3, 2014. The letter requested that they discontinue distribution and quarantine the products and return the products to Teleflex. The letter also requested a sub-recall.
Recalling firm
- Firm
- Teleflex Medical
- Address
- 4024 Stirrup Creek Dr, Durham, North Carolina 27703-9000
Distribution
- Distribution pattern
- Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, WY, and the countries of United Arab Emirates, Australia, Belgium, Brazil, Canada, Chile, China, France, Great Britain, Japan, India, Indonesia, Israel, South Korea, Mexico, Philippines, Saudi Arabia, Singapore, Taiwan, Thailand, Uruguay, and Zambia.
Timeline
- Recall initiated
- 2014-09-03
- Posted by FDA
- 2014-12-12
- Terminated
- 2017-08-14
- Status
- —