Recalls / —
—#131612
Product
PKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide: Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Venezuela
- FDA product code
- KOG
- Affected lot / code info
- Serial/Lot 0915269/0817086 1015777 1016562/1015776 111333/1015777 111333-1015777
Why it was recalled
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Root cause (FDA determination)
Error in labeling
Action the firm took
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Timeline
- Recall initiated
- 2014-11-17
- Posted by FDA
- 2014-12-12
- Terminated
- 2015-07-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #131612. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.