FDA Device Recalls

Recalls /

#131646

Product

PKG, RING HANDLE, P/N 0250080236. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

FDA product code
KOG
Affected lot / code info
514021 614128 717447 814171 816251 816252 913442 913545 913581 913811 913812 913984 913985 914008 914371 914372 914894 1016560 1016736 1016739 1016846 1016847 1113330 1113331 1113332 1114831 1116462 1116463 1216409 1314473 1314474 1314920 1314924 1314925 1315268 1315404 1316600 1316601 1316894 1316895 1316896 1410871 1410872 1410873 1410874 1411128

Why it was recalled

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Root cause (FDA determination)

Error in labeling

Action the firm took

Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Timeline

Recall initiated
2014-11-17
Posted by FDA
2014-12-12
Terminated
2015-07-17
Status