FDA Device Recalls

Recalls /

#131676

Product

PKG, MIXTER FORCEPS, P/N 0250080285. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

FDA product code
KOG
Affected lot / code info
41350 44269 442637 647415 743732 845623 846466 846767 847691 940788 941241 941606 942276 942920 943500 943533 945623 1040414 1041622 1042249 1042620 1042673 1042674 1042675 1046593 1047021 1140650 1142620 1143129 1144315 1144939 1145976 1146830 1240976 1242463 1244169 1246128 1246649 1341525 1341993 1342760 1343129 1346546 1346648 1440442 1442804

Why it was recalled

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Root cause (FDA determination)

Error in labeling

Action the firm took

Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Timeline

Recall initiated
2014-11-17
Posted by FDA
2014-12-12
Terminated
2015-07-17
Status