#131795

Product

PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

FDA product code
KOG
Affected lot / code info
140289 641421 840358 848483 940108 940131 946000 1040117 1040987 1041626 1042737 1042745 1045058 1046928 1143580 1145635 1145811 1145930 1146540 1241586 1242142 1242643 1243443 1243554 1244523 1340562 1341469 1341471 1341995 1343407 1343706 1344781 1346104 1346430 1346651 1346652

Why it was recalled

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Root cause (FDA determination)

Error in labeling

Action the firm took

Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Timeline

Recall initiated
2014-11-17
Posted by FDA
2014-12-12
Terminated
2015-07-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #131795. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting · FDA Device Recalls