FDA Device Recalls

Recalls /

#131796

Product

PKG, 5MM ROTATING HANDLE, INSULATED SHAFT/ HANDLE 45CM LENGTH, P/N 0250080595 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

FDA product code
KOG
Affected lot / code info
142343 0200310C 0200311C 0342702C 0642220D 0647222D 0744685D 0744999D 0745480D 0745846D 0845031D 0845499D 0941244D 0942148D 0943252D 0943439D 0943710D 0944124D 0944420D 0944694D 1040118D 1040580D 1040613D 1042491D 104249ID 1042728D 1042734D 1042746D 1045483D 1141563D 1143001D 1143497D 1144643D 1144931D 1146202D 1240141D 1241177D 1243306D 1243442 1243553 1243553D 1244642 1244643 1244701D 1244706 1245114 1245542 1245542D 1245595 1246201 1246201D 13039742 1343018 1343414 1344033 1344033D 1345422 1345422D 1345738 1345738D 1346232 1441637 1442630 352499

Why it was recalled

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Root cause (FDA determination)

Error in labeling

Action the firm took

Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Timeline

Recall initiated
2014-11-17
Posted by FDA
2014-12-12
Terminated
2015-07-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #131796. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.