Recalls / —
—#131814
Product
PKG, 5MM INSERT, ENDO GRASPER, 45CM, P/N 0250080726 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
- FDA product code
- KOG
- Affected lot / code info
- 143498 143553 1044398 1044399 1146774 1343552 1344884 1345231 1345234 1345414 1346544 1346661 1440959
Why it was recalled
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Root cause (FDA determination)
Error in labeling
Action the firm took
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Timeline
- Recall initiated
- 2014-11-17
- Posted by FDA
- 2014-12-12
- Terminated
- 2015-07-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #131814. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.