Product
PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 20CM, P/N 0250282000 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
- FDA product code
- KOG
- Affected lot / code info
- 342970 645484 743722 747490 747891 840091 945208 945712 946001 1040121 1041472 1042482 1042952 1046585 1140928 1143074 1143076 1143121 1145963 1146536 1241511 1244708 1246167 1341512 1343689 1344271 1344682 1440414
Why it was recalled
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Root cause (FDA determination)
Error in labeling
Action the firm took
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Timeline
- Recall initiated
- 2014-11-17
- Posted by FDA
- 2014-12-12
- Terminated
- 2015-07-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #131844. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.