FDA Device Recalls

Recalls /

#132123

Product

Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer

FDA product code
CGAGlucose Oxidase, Glucose
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K032311
Affected lot / code info
Model/Catalog/Part #: 1. cobas b221 <6>=Roche OMNI S6 SYSTEM; and 2. 03337154692 - cobas b 221<6>=Roche OMNI S6 SYSTEM CU Potassium - CEM, Calcium - JFP, Sodium - JGS, Chloride - CGZ, Blood Gases (PCO2, PO2) and pH - CHL, Hemoglobin -GKR, Hematocrit - GKF, Oxygen saturation, Carboxyhemoglobin - GHS, Oxyhemoglobin - GGZ, Methemoglobin - KHG, Deoxyhemoglobin - GKA, Lactic Acid - KHP, Glucose- CGA, Urea Nitrogen - CDS.

Why it was recalled

Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected.

Root cause (FDA determination)

Material/Component Contamination

Action the firm took

Roche Diagnostics has confirmed an issue with the calibration and QC stability of MSS parameters for cobas b 221 <6> systems. Internal investigations identified the formation of white deposits in the fluidic system, especially in the SD cartridge tubing, as the cause for the premature failure of the MSS parameters. These white deposits can potentially affect the fluidic pathways of both the Standby solution and the calibration solutions Cal1 to Cal4 (from the S3 Fluid Pack) and may reduce the lifetime of metabolite sensor cartridges. The cause of the white deposit has been identified as a chemical component of the S3 Fluid Pack solutions, as well as environmental bacterial contamination. Since this issue involves the S3 Fluid Pack, only the cobas b 221 <6> system is impacted. This issue may have an effect on the calibration of Glucose and Lactate parameters, which may lead to patient results being affected, as calibration drifts may not be detected by calibration and QC procedures. Glucose and Lactate are the only parameters affected by this issue. As a result, Roche is implementing an interim corrective action to be performed on your systems on a regular basis by Roche Field Engineering Specialist (FESs). Read the Urgent Medical Device Correction (UMDC) in its entirety. " Continue to run and maintain your cobas b 221 <6> system as per the Instructions for Use. " Complete the attached fax form and fax it to 1-888-656-6385. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC. Public Contact is Todd Siesky, at todd.siesky@roche.com.

Recalling firm

Firm
Roche Diagnostics Operations, Inc.
Address
9115 Hague Rd, Indianapolis, Indiana 46256-1025

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2014-12-18
Posted by FDA
2015-02-02
Terminated
2017-02-06
Status

Source: openFDA Device Recall endpoint. Recall record ID #132123. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.