Product

Fixed Tibial bearing ATTUNE INTUITION Impactor The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
CATALOG NO. : ATTUNE INTUITION Impactors  Fixed Tibial Bearing, (254401003)  Barcode GTIN Number:  Fixed Tibial Bearing, ......... 10603295130192   ALL LOTS

Why it was recalled

ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.

Root cause (FDA determination)

Device Design

Action the firm took

DePuy Synthes sent an Urgent Information letter dated November 10, 2014, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to complete the Hospital/User Facility Acknowledgement Car and return to their Deputy Orthopaedics Sales Consultant or fax to 574-371-4939. Customer swere also asked to retain a copy of the complete Acknnowlegement Card in their files along with the notice. Customers with questions were instructed to call 574-371-4917 (M-F; 8 a.m. 5 p.m. EDT). For questions regarding this recall cal 574-371-4917.

Recalling firm

Firm
DePuy Orthopaedics, Inc.
Address
700 Orthopaedic Dr, Warsaw, Indiana 46582-3994

Distribution

Distribution pattern
Worldwide Distribution: US (nationwide) including P.R. except AK and RI; and Internationally to: Chile, Canada, Australia, Austria, Belgium, China, France, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK).

Timeline

Recall initiated
2014-11-10
Posted by FDA
2015-01-05
Terminated
2015-12-08
Status

Source: openFDA Device Recall endpoint. Recall record ID #132180. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.