FDA Device Recalls

Recalls /

#133894

Product

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

FDA product code
DWCController, Temperature, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K071521
Affected lot / code info
HX2 Temperature Management System, 120V  Catalog # 809810  serial numbers: 00016, 0001 1116  HX2 Temperature Management System, 100V Catalog # 812231  serial numbers: 0001, 1002, PT0003, PT0004   HX2 Temperature Management System, 240V Catalog # 812240  serial numbers: 0001, 0002  HX2 Temperature Management System, Trial Kit Catalog # 809810TK  serial numbers: 1001 1042

Why it was recalled

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Root cause (FDA determination)

Device Design

Action the firm took

On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.

Recalling firm

Firm
Terumo Cardiovascular Systems Corporation
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.

Timeline

Recall initiated
2015-02-06
Posted by FDA
2015-02-13
Terminated
2016-10-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #133894. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.