FDA Device Recalls

Recalls /

#134235

Product

Vitek 2 AST-P635 REF 416 911 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae.

FDA product code
LTWSusceptibility Test Cards, Antimicrobial
Device class
Class 2
Medical specialty
Microbiology
Affected lot / code info
lot 735352820, expiration 11AUG16; lot 735348620, expiration 30JUN16; lot 735346520, expiration 09JUN16; lot 735342320; expiration 28APR16; lot 735318020, expiration 29AUG15; lot 735315920, expiration 08AUG15; lot 735323710, expiration 25OCT15; lot 735310210, expiration 12JUN15

Why it was recalled

The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.

Root cause (FDA determination)

Employee error

Action the firm took

Biomerieux, Incs sent an Urgent Product Correction Notice on February 25, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to do the following: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " For the referenced test kits (VITEK 2 ASTP635, ASTP640 or ASTP641), review the claimed organisms in the Package Insert antibiotic table to confirm which organisms can be tested. " Refrain from using the referenced test kits to determine the susceptibility of Streptococcus pneumoniae. " Reference your Laboratory Standard Operating Procedures to identify card type used for specific organisms. " Please store this letter with your bioMrieux instrument documentation. Please complete the attached Acknowledgement Form and return it to bioMrieux, Inc. Customers with questions were instructed to contact their local BioMerieux representative. For questions regarding this recall call 314-731-8526.

Recalling firm

Firm
Biomerieux Inc
Address
595 Anglum Rd, Hazelwood, Missouri 63042-2320

Distribution

Distribution pattern
Worldwide Distribution to United Kingdom and Turkey. NO U.S. distribution.

Timeline

Recall initiated
2015-02-25
Posted by FDA
2015-04-08
Terminated
2017-03-22
Status

Source: openFDA Device Recall endpoint. Recall record ID #134235. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.