FDA Device Recalls

Recalls /

#134707

Product

da Vinci S Surgical System IS2000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.

FDA product code
KKXDrape, Surgical
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K050322
Affected lot / code info
Product number 420256-01; da Vinci S Disposable Accessory Kit, 3 ARM, 5 Pack.  Lot Numbers for Cloudy/Waxy issue: ALL lot numbers  Lot Numbers numbers for Tear issue: D142205, D142275, D142345, D142375, D142625, D142695, D142765, D142835, D142905, D142975, D143035, D143115, D143175, D143255, D143315, D143385, D143445, D150085, D150215, D150295, D150365, D150435, D150515, D150575, D150655

Why it was recalled

Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.

Root cause (FDA determination)

Packaging process control

Action the firm took

Firm issued field safety notice to all customers on March 16, 2015. Field safety notice includes acknowledgement form to be returned to recalling firm. A second letter was sent on May 12, 2015 to inform users that the date that the drape product unaffected by the issue will not be availalble untisl October/November 2015, not May 2015 as stated in the original communication.

Recalling firm

Firm
Intuitive Surgical, Inc.
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, California 94086-5304

Distribution

Distribution pattern
Worldwide Distribution-Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China (including Hong Kong), Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, France, Germany, Greece, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Mauritius, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam.

Timeline

Recall initiated
2015-03-16
Posted by FDA
2015-04-06
Terminated
2015-12-03
Status

Source: openFDA Device Recall endpoint. Recall record ID #134707. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.