FDA Device Recalls

Recalls /

#134907

Product

Edwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx Only, Manufacturer Edwards Lifesciences LLC, Irvine, CA.-Model Numbers: FEMII008A, FEMII008AT, FEMII008V, FEMII010A, FEMII010AT, FEMII010V, FEMII012A, FEMII012AT, and FEMII012V. Edwards Femoral Access Cannulae are intended to provide a means of draining the blood flow (venous), or perfusing blood into the body (arterial) of a patient during cardiopulmonary bypass procedures.

FDA product code
DWFCatheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K140208
Affected lot / code info
Model/Lot Numbers: FEMII008A/59751073, 59775775, 59775776, 59775777 & 59852930; FEMII008AT/59807985, 59867050, & 59873263;  FEMII008V/59873250 & 59873251; FEMII010A/59740468, 59773806, 59792415, 59792416 & 59852934;  FEMII010AT/ 59747819, 59807986,59852935, 59890916 & 59896910; FEMII010V/59751074, 59849119 & 59890924; FEMII012A/ 59801792,59867064, 59884766 & 59884778  FEMII012AT/59852940 & 59867051; and  FEMII012V/59723307, 59796683, 59849124 & 59873252

Why it was recalled

Edwards Lifesciences is recalling Fem-Flex II Pediatric Femoral Arterial Cannula sizes 8, 10, 12 French because of the potential of tissue damage caused by a protruding wire located at the tip of the cannula.

Root cause (FDA determination)

Process change control

Action the firm took

The firm, Edwards, sent an "URGENT - PRODUCT RECALL - ACTION REQUIRED" letter dated March 23, 2015 via FED EX on March 27, 2015 to their customers. The letter described the product, problem, and actions to be taken. The customers were instructed to review entire inventory for the lots listed; complete and return attached acknowledgment form via fax to Edwards Customer Service at 800.422.9329. within three days of receipt of this Field Safety Notice; contact Customer Service at 800.422.3278 to obtain an RGA number and replacement product; return affected product to Edwards Lifesciences, Attn: Cirilo Chaparro, 12050 Lone Peak Drive, Draper, UT 84020, Attention: RECALL, RGA#XXX; and transfer this notice to other organizations if the affected devices have been transferred to any another facilities. If you have any questions that have not been answered by this letter, please call Edwards Customer Services at 800.424.3278 from the hours of 6:00AM - 4:30PM PST; Edwards Customer Service at (800) 268-3993 from 8:00AM 4:30PM Eastern Time or contact your Edwards sales representative concerning the recall.

Recalling firm

Firm
Edwards Lifesciences, LLC
Address
12050 Lone Peak Pkwy, Draper, Utah 84020-9414

Distribution

Distribution pattern
Nationwide Distribution.

Timeline

Recall initiated
2015-03-23
Posted by FDA
2015-04-02
Terminated
2017-04-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #134907. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.