Product
2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. - Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
- FDA product code
- KDI — Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K080964, K113427, K120017, K120505, K121341
- Affected lot / code info
- Serial Numbers: 0T0S100031- 101358, 1T0S101359- 127938, 2T0S109489- 118130, 3T0S118131- 126163, 4T0S126164- 135408, 9T0S100002, 9T0S100004, 9T0S100007- 100010, 9T0S100012, 9T0S100014- 100016
Why it was recalled
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Root cause (FDA determination)
Software design
Action the firm took
The firm, Fresenius Medical Care, sent an "URGENT - MEDICAL DEVICE CORRECTION" letter dated February 19, 2015 to its Consignees via certified mail on February 26, 2015. The letter describes the product, problem and actions to be taken. The consignees were instructed to sign and return the attached "Reply Form" via fax to: 781-699-9769 Attn: Quality Department or Scan and E-mail to: NOTIFYRA@fmc-na.com to confirm that they received and understand the customer notification. The firm will be providing a software update and stated that in the near future, your local clinic will be contacting you about installing the software update. If you have any questions pertaining to this Customer Notification, please contact your local clinic, 800-227-2572 or the Corrections and Removals Manager at 800-662-1237.
Recalling firm
- Firm
- Fresenius Medical Care Holdings, Inc.
- Address
- 920 Winter St, Waltham, Massachusetts 02451-1521
Distribution
- Distribution pattern
- Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Timeline
- Recall initiated
- 2015-02-20
- Posted by FDA
- 2015-04-29
- Terminated
- 2021-05-18
- Status
- —