Product
Lifesaver Single Patient Use Manual Resuscitator Product Usage: The Hudson RCI Lifesaver Single Patient Use manual resuscitator with pressure monitoring port is a disposable medical device intended for use on patients requiring temporary augmentation of ventilation with or without supplemental oxygen delivery during episodes of acute ventilatory failure or insufficiency.
- FDA product code
- BTM — Ventilator, Emergency, Manual (Resuscitator)
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K964719
- Affected lot / code info
- Product Code 5361 - 6 Digit Lot No. 140514 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140616 4 Digit Lot No. 1425, 6 Digit Lot No. 140617 4 Digit Lot No. 1425; 6 Digit Lot No. 140621 4 Digit Lot No. 1425; 6 Digit Lot No. 140806 4 Digit Lot No. 1432; 6 Digit Lot No. 141103 4 Digit Lot No. 1445; 6 Digit Lot 141117 - No. 4 Digit Lot No. 1447; 6 Digit Lot No. 141201- 4 Digit Lot No. 1449; 6 Digit Lot No. 141224 - 4 Digit Lot No. 1452; 6 Digit Lot No. 141227 - 4 Digit Lot No. 1452; 6 Digit Lot No. 150112 - 4 Digit Lot No. 1503; Product Code 5362 - 6 Digit Lot No. 140504 - 4 Digit Lot No. 1419; 6 Digit Lot No. 140515 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140517 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140617 - 4 Digit Lot No. 1425; 6 Digit Lot No. 140621 - 4 Digit Lot 1425; No. 6 Digit Lot No. 140812 - 4 Digit Lot No. 1433; 6 Digit Lot No. 140819 - 4 Digit Lot No. 1434; 6 Digit Lot No. 140823 - 4 Digit Lot No. 1434; 6 Digit Lot No. 140929 - 4 Digit Lot No. 1440; 6 Digit Lot No. 141110 - 4 Digit Lot No. 1446; 6 Digit Lot No. 141115 - 4 Digit Lot No. 1446; 6 Digit Lot No. 141201 - 4 Digit Lot No. 1449; 6 Digit Lot No. 141214 - 4 Digit Lot No.1451; 6 Digit Lot No. 150112 4 Digit Lot No. 1503; Product Code 5364 6 Digit Lot No. 140507 - 4 Digit Lot No. 1419; 6 Digit Lot No. 140508 4 Digit Lot No. 1419; 6 Digit Lot No. 140514 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140608 - 4 Digit Lot No. 1424; 6 Digit Lot No. 140614 - 4 Digit Lot No. 1424; 6 Digit Lot No. 140623 - 4 Digit Lot No. 1426; 6 Digit Lot No. 14628 - 4 Digit Lot 1426 - No. 6 Digit Lot No. 140714 - 4 Digit Lot No. 1429; 6 Digit Lot No. 140719 - 4 Digit Lot No. 1429; 6 Digit Lot No. 14806 - 4 Digit Lot No. 1432; 6 Digit Lot No. 140809 4 Digit Lot No. 1432; Product Code 5466 6 Digit Lot No. 140520 - 4 Digit Lot No. 1421; 6 Digit Lot No. 140923 - 4 Digit Lot No. 1439; 6 Digit Lot No. 140924 - 4 Digit Lot No. 1439; Product Code 45362 - 6 Digit Lot No. 140317 - 4 Digit Lot No. 1412; 6 Digit Lot No. 140630 - 4 Digit Lot No. 1427; 6 Digit Lot No. 141013 - 4 Digit Lot No. 1442.
Why it was recalled
The intake port may be blocked which can cause the bag to fail to fill.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
Teleflex Medical sent an Urgent Medical Device Recall Notification letter dated May 14, 2015, to all consignees asking them to immediately discontinue distribution and quarantine the recalled product. The letter also requested a sub-recall of the product. The recall letter also included a response form which is to be returned to Teleflex by emailing it to recall@teleflex.com or fax it to 1-855-419-8507, Attn: Customer Service. Customers with questions were instructed to contact their local sales representative or Customer Service at 1-866-246-6990. For questions regarding this recall call 610-378-0131.
Recalling firm
- Firm
- Teleflex Medical
- Address
- 2917 Weck Dr, Research Triangle Park, North Carolina 27709-0186
Distribution
- Distribution pattern
- Worldwide Distribution - US Nationwide in the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IA, KS, LA, MA, MI, MN, MS, MO, NE, NH, NY, NC, OH, OK, PA, RI, SD, TN, TX, VA, WA, including Puerto Rico and the countries of Australia, Bahamas, Canada, Guatemala, and Mexico.
Timeline
- Recall initiated
- 2015-05-14
- Posted by FDA
- 2015-06-24
- Terminated
- 2024-05-13
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #137056. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.