FDA Device Recalls

Recalls /

#137176

Product

Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.

FDA product code
HAWNeurological Stereotaxic Instrument
Device class
Class 2
Medical specialty
Neurology
510(k) numbers
K022365
Affected lot / code info
Serial numbers 100019, 100984, 100965, 100295, 100979, 100773, 100946, 100797, 100947, 100956, 100966, 100649, 100967, 100944, 100941, 100945, 100970, 100950, 100976, 100951, 100728, 100952, 100248, 100953, 100641, 100954, 100463, 100955, 100981, 100957, 100978, 100958, 100106, 100959, 100939, 100960, 100973, 100961, 100236, 100962, 100237, 100963, 100417, 100964. 100940, 100968, 100975, 100969, 100977, 100972, 100105, 100974, 100214, 100980, 100492, 100982, 100493, 100983, 100948, 100986, 100704, 100987, 100949, 100988, 100227, 100989, 100971, 100133, 100134, 100942.

Why it was recalled

Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.

Root cause (FDA determination)

Process control

Action the firm took

Stryker sent an Urgent Medical Device Recall Notification letter dated May 18, 2013 to Domestic customers notice via FedEx overnight and Sales Representatives via email. The International Subsidiaries will also be notified via email. Customers are instructed to immediately review the Recall Notification adn complete the following: Immediately check all stock areas and/or operating room storage for affected equipment found. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many affected items are currently in inventory. Please complete and return the BRF even if they dont have any affected product on hand. If customers have further distributed this product, they are asked to forward this letter and the attached BRF to all affected locations. Please indicate each location and serial number(s) on the BRF. Fax (866-521-2762) or email (kellyjo.davis@stryker.com) the completed BRF to Stryker Instruments Regulatory Department, Attn: Kelly Jo Davis. A Stryker Representative will contact the facility to set up a time to perform the additional inspection of the weld seam on the Navigation System II-Cart, Articulated Arm Camera.

Recalling firm

Firm
Stryker Instruments Div. of Stryker Corporation
Address
4100 E Milham Ave, Portage, Michigan 49002-9704

Distribution

Distribution pattern
Worldwide Distribution - US Nationwide in the states of AR, CO, FL, GA, IN, KS, KY, MA, MI, NC, NE, NJ, NY, OH, OK, PA, TX, WA, WY and the countries of Canada, Europe, and Asia.

Timeline

Recall initiated
2015-05-04
Posted by FDA
2015-06-16
Terminated
2016-05-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #137176. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.