#137248

Product

SOMATOM Emotion 16, the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K133424
Affected lot / code info
SOMATOM Emotion 16 (2007) 10165977  with serial numbers:  86514 32490 80553 32552 32530 86406 86558 80542 80461 86405 32512 80441 86431 80467 86425 80442 86429 86557 80434 32581 32637 32597 80448 80431 32624 32469 80446 32578 80430

Why it was recalled

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Root cause (FDA determination)

Software design

Action the firm took

Siemens sent an Customer Safety Advisory Notice dated April 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A remote software update is scheduled for release that will resolve the issue. For further questions, please call (610) 219-6300.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355-1406

Distribution

Distribution pattern
Nationwide Distribution including Puerto Rico.

Timeline

Recall initiated
2015-04-20
Posted by FDA
2015-05-22
Terminated
2016-09-01
Status

Source: openFDA Device Recall endpoint. Recall record ID #137248. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.