Recalls / —
—#137351
Product
Medtronic Navigated Cannulated Taps Individual cannulated taps are packaged in a clear plastic tube with protective plastic end caps. Kit/sets contain various devices including the taps. The kit/set configuration varies dependent on the part number. The devices are assembled and packaged into a shipping carton or into a holding tray which is then placed into a shipping carton. Designed for use in spine surgery to assist in the tapping procedure and the placement of screws during the surgery.
- FDA product code
- HAW — Neurological Stereotaxic Instrument
- Device class
- Class 2
- Medical specialty
- Neurology
- Affected lot / code info
- Correction includes all lots of products distributed between January 2007 and March 2015. P/N: 9734240,Nav 6.5mm Tap (Cannulated),9734239, Nav 5.5mm Tap (Cannulated), 9734238,Nav 4.5mm Tap (Cannulated), 9734302,6.5mm Tap (Cannulated),9734300,5.5mm Tap (Cannulated),9734298,4.5mm Tap (Cannulated),9734303,7.5mm Tap (Cannulated),9734301,6.0mm Tap (Cannulated),9734299,5.0mm Tap (Cannulated),9734304,8.5mm Tap (Cannulated),9730944,4.5 mm Cannulated Tap,9730945,5.5 mm Cannulated Tap,9730946,6.5 mm Cannulated Tap,9733182,4.0 mm Cannulated Tap, 1.86P,9733183,4.5 mm Cannulated Tap,9733184,4 mm Cannulated Tap, 1.5P, UCSS,960-347,7.5 Old Round APT (Cannulated Tap-Tip only),960-345,6.5 Old Round APT (Cannulated Tap-Tip only),960-344,5.5 Old Round APT (Cannulated Tap-Tip only) The above referenced cannulated taps are included in the kits and,9731921,KIT 9731921 INSTRUMENT SPINE BASIC,9735101,INST KIT 9735101 TAP ASM 4.5 G5, 9735102,INST KIT 9735102 TAP ASM 5.5 G5,9735103,INST KIT 9735103 TAP ASM 6.5 G5,9735107,INST KIT 9735107 TAP ASM 5.0 G5,9735108,INST KIT 9735108 TAP ASM 6.0 G5,9735109,INST KIT 9735109 TAP ASM 7.5 G5,9735149,INST KIT 9735149 CANNULATED 6.5MM TAP,9735150,INST KIT 9735150 CANNULATED 5.5MM TAP,9735151,INST KIT 9735151 CANNULATED 4.5MM TAP,9735168,INST KIT 9735168 4MM TAP 1.86P,9735169,INST KIT 9735169 4.5MM CANN TAP,9735170,INST KIT 9735170 4MM TAP 1.5P,9734490,INST SET 9734490 LEGACY,9734491,INST SET 9734491 SOLERA STD,9734492,INST SET 9734492 SOLERA DEF UPGRADE,9735469,TAPS 9735469 CD HORIZON SOLERA,9734631,TD SET 9734631 LEGACY STD TAPS/DRIVERS,9734632,TD SET 9734632 SOLERA STD TAPS/DRIVERS,9733370,VERTEX MAX SET, 9733370, NAVIGATED,9734838,KIT 9734838 SOLERA 8.5 CANNULATED TAP,
Why it was recalled
Medtronic Navigation is initiating a field correction due to the potential for injury which could occur as a result of cleaning the Medtronic Navigated Cannulated Taps.
Root cause (FDA determination)
Error in labeling
Action the firm took
Consignees were notified via Urgent Medical Device Correction Notification on 5/26/15. The letter identified the affected product and described the issue. This field correction does not require the return of the subject products. Revised instructions for use are being provided to the identified consignees.
Recalling firm
- Firm
- Medtronic Navigation, Inc.
- Address
- 826 Coal Creek Cir, Louisville, Colorado 80027-9710
Distribution
- Distribution pattern
- Worldwide Distribution.
Timeline
- Recall initiated
- 2015-05-26
- Posted by FDA
- 2015-06-26
- Terminated
- 2015-10-23
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #137351. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.