FDA Device Recalls

Recalls /

#137730

Product

VersaTREK Instrument Series 528 REF 6528, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.

FDA product code
MDBSystem, Blood Culturing
Device class
Class 1
Medical specialty
Microbiology
Affected lot / code info
Serial numbers: 0169050501141020, 0169059801141031, 0169064701141210, 0169066801150105, 2200X1113, 2215X1113, 2236X1213, 2244X1213, 2272X0214, 2273X0214, 2274X0214, 2275X0214, 2298X0714, 2299X0714, 2300X0714, 2301X0714, 2303X0714, 2304X0714, 2326X0814, 2327X0814, and 2328X0814;

Why it was recalled

Use of the recalled product may result in false positive reports.

Root cause (FDA determination)

Release of Material/Component prior to receiving test results

Action the firm took

Remel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of the notice of the affected product. If product was further distributed customers were instructed to contact those entities, advise them of the situation and provide them with a copy of the letter. Customers with questions were instructed to contact the Technical Services Department at 800-642-7029. For questions regarding this recall call 770-409-0713.

Recalling firm

Firm
Remel Inc
Address
12076 Santa Fe Trail Dr, Lenexa, Kansas 66215-3519

Distribution

Distribution pattern
Worldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingdom, Uruguay, and Vietnam.

Timeline

Recall initiated
2015-05-19
Posted by FDA
2015-06-25
Terminated
2017-03-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #137730. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.