FDA Device Recalls

Recalls /

#137735

Product

ATTUNE CONV FB CR TB TRL SZ3 - 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.

FDA product code
MBHProsthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K140881
Affected lot / code info
Product Codes:254501963, 254501964, 254501965, 254501966, 254501967, 254501968, 254501969, and 254501970. GTIN: 10603295423355, 10603295423362, 10603295423379, 10603295423386, 10603295423393, 10603295423409, 10603295423416, and 10603295423423.

Why it was recalled

Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

Root cause (FDA determination)

Device Design

Action the firm took

URGENT INFORMATION: MEDICAL DEVICE CORRECTION was issued for all lots of the affected devices. The device correction is to inform US Distributors and Medical Professionals about the issue, how to prevent the issue, and any actions to take. US Distributors will provide Hospitals/User Facilities with the Medical Professional Notice and Acknowledgment Cards. US Distributors will also work with the sales consultants to ensure all consultants complete sales training regarding this issue.

Recalling firm

Firm
DePuy Orthopaedics, Inc.
Address
700 Orthopaedic Dr, Warsaw, Indiana 46582-3994

Distribution

Distribution pattern
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to CANADA , AUSTRALIA, AUSTRIA, BELGIUM, ENGLAND/ UK, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JAPAN, MALAYSIA, NETHERLANDS, NEW ZEALAND, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN & IBERIA, SWEDEN, SWITZERLAND, THAILAND, UAE, CHILE, COLOMBIA, MEXICO, PERU, and PUERTO RICO.

Timeline

Recall initiated
2015-06-12
Posted by FDA
2015-07-20
Terminated
2015-11-12
Status

Source: openFDA Device Recall endpoint. Recall record ID #137735. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.