Recalls / —
—#137950
Product
Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element insertion. Manual orthopedic surgical instrument.
- FDA product code
- HSB — Rod, Fixation, Intramedullary And Accessories
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K131548
- Affected lot / code info
- Part Number Lot Numbers 03.037.024 T102759
Why it was recalled
The height of one of the three guiding pins of the Helical Blade Inserter for the TFNA is oversized. The Helical Blade Inserter is therefore unable to pass through the Blade/Screw Guide Sleeve (part 03.037.017).
Root cause (FDA determination)
Process control
Action the firm took
A recall notification letter, dated June 10, 2015, was sent to sales consultants and Synthes distribution sites requesting return of affected devices.
Recalling firm
- Firm
- Synthes (USA) Products LLC
- Address
- 1301 Goshen Pkwy, West Chester, Pennsylvania 19380-5986
Distribution
- Distribution pattern
- Nationwide Distribution.
Timeline
- Recall initiated
- 2015-06-10
- Posted by FDA
- 2015-07-16
- Terminated
- 2016-08-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #137950. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.