Product

A.L.P.S. Foot Locking Calcaneus Plate,Small-Right. Provides the orthopaedic surgeon a means of bone fixation and helps generally in the management of fractures and reconstructive surgeries.

FDA product code
HRSPlate, Fixation, Bone
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K093474, K111663
Affected lot / code info
Part Number: 816209001; Lots: 015840, 124980, 948210 and P0059, P0076, or P0061 will be etched on the plate itself.

Why it was recalled

The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the plate if it does not achieve proper purchase. This may require a revision surgery to replace or remove the screw and/or plate. The threads, not uniformly loaded, may shear off, causing metal slivers.

Root cause (FDA determination)

Process change control

Action the firm took

On June 4, 2015 an URGENT MEDICAL DEVICE REMOVAL letter was sent to all consignees. This action requires the immediate location and discontinued use of the product and its return to Biomet. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation; Carefully follow the instructions on the enclosed "Response Form"; Email a copy of the Response Form to audrey.daenzer@biomet.com prior to return of product; Use priority carrier for your shipment; and If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear Risk/Recall Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products. Please confirm receipt of this notice by sending back the response form within three (3) business days. Thank you in advance for your assistance and prompt attention. On behalf of Biomet, I apologize for any inconvenience this may cause. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.

Recalling firm

Firm
Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582-6989

Distribution

Distribution pattern
Distributed in the states of TX, MD, PA, NJ, NC, MA, AZ, GA, FL, and AR. and the countries of Costa Rica CP and The Netherlands.

Timeline

Recall initiated
2015-06-04
Posted by FDA
2015-07-08
Terminated
2015-12-22
Status