FDA Device Recalls

Recalls /

#138147

Product

Ring Transjugular Intrahepatic Access Set Product Usage: The Ring Transjugular Intrahepatic Access Sets are intended for transjugular liver access in diagnostic and interventional procedures. The Ring Transjugular Intrahepatic Access Set consists of the following components: Flexor RTPS Guiding Sheath, needle, catheter, Dilator, Torcon NB Advantage Catheter, Van Andel Dilatation Catheter, Curved Newton Wire Guide, Amplatz Extra Stiff Wire Guide.

FDA product code
GBZCatheter, Cholangiography
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Product Number: RTPS-100-10.0, Catalog Number G29769

Why it was recalled

Cook Inc. is initiating a voluntary recall of the Ring Transjugular Intrahepatic Access Set and the Flexor RTPS Introduce Sheath due to complaints of the dilators being too long.

Root cause (FDA determination)

Component change control

Action the firm took

Written recall communication letters were shipped via FedEx on June 05, 2015 to all US Consignees. Customers were asked to complete and return response forms and recalled products. Recalled devices will be held for final disposition and destruction. Customers wree asked to forward notices if recalled devices were further distributed. Effectivenes checks will be conducted by mail, phone and visits. Customers may contact Cook Medical Relations (800) 457-4500 or 1 (812)- 339-2235 Monday through Friday from 7:30 AM to 5:00 PM or e-mail at CustomerrelationsNA@cook medical.com.

Recalling firm

Firm
Cook Inc.
Address
750 N Daniels Way, Bloomington, Indiana 47404-9120

Distribution

Distribution pattern
Worldwide Distribution - US Nationwide in the state of: AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada and EU.

Timeline

Recall initiated
2015-05-28
Posted by FDA
2015-07-18
Terminated
2017-05-30
Status