Recalls / —
—#138201
Product
GE Healthcare Discovery IGS 730. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional and image-guided surgical procedures.
- FDA product code
- IZI — System, X-Ray, Angiographic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K122457
- Affected lot / code info
- Serial Number System ID 0000012C2A0003 214345IGS730 0000015A2A0012 AM137CZ18 0000015A2A0008 210060RX14 15A2A0011 GON4289057
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter on 5/29/2015- GEHC Ref#12231- dated May 29, 2015 to their Consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and contact Information. The customers were instructed that you may continue to use the system and avoid rapid direction changes with the gantry motions and prevent potential collisions with other equipment. If the failure occurs during interventional examination, please ensure that you have established procedures for handling patients in case of the loss of gantry motion as stated in your product labeling. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recalling firm
- Firm
- GE Healthcare
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, GA, IL, IN, KS, MS, NV, OH, OK, OR, PA, TN, TX, VA, WA and WI; and countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAPAN, RUSSIA, UNITED KINGDOM, and VIETNAM.
Timeline
- Recall initiated
- 2015-05-29
- Posted by FDA
- 2015-07-03
- Terminated
- 2015-10-28
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #138201. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.