FDA Device Recalls

Recalls /

#138723

Product

rHead Lateral Stem, Size 1, Do Not Reuse, Sterile. Intended for replacement of the proximal end of the radius.

FDA product code
JDBProsthesis, Elbow, Semi-Constrained, Cemented
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K102180
Affected lot / code info
Part Number: 310-0000 (Lateral Assembly, Radial Implant, Size 1) Lot #'s 22288301, 22890701, 23424401, 23916901, 19058-S2, 19452-S2, 19661-S2, 19662-S2, 19712-S2, 19736-S2 and 19841-S2; Part Number: 310-0001 (Lateral Assembly, Radial Implant, Size 2) Lot #'s 22629201, 22890801, 23424501, 23917001, 19453-S2, 19737-S2, 19765-S2, 19766-S2 and 19793-S2; Part Number:310-0002 (Lateral Assembly, Radial Stem Implant, Size 3) Lot #'s 22288401, 22629301, 23198901, 23434601, 23917101, 19678-S2, 19767-S2, 19768-S2, 19794-S2 and 19842-S2; Part Number: 310-0003 (Lateral Assembly, Radial Implant, Size 4) Lot #'s 22288501, 23199001, 23424701, 23917201, 19068-S2, 19679-S2, 19713-S2, 19738-S2, 19795-S2 and 19843-S2; Part Number: 310-0004 (Lat Assembly, Rad Stem Implant, Collar 6mm, Size1) Lot #'s 22288601, 23424801, 19057-S2, 19067-S2, 19454-S2, 19739-S2, 19796-S2 and 19844-S2; Part Number: 310-0005 (Lat Assembly, Rad Stem Implant, Collar 6mm, Size 2) Lot #'s 23424901, 23917301, 19063-S2, 19124-S2, 19455-S2, 19714-S2, 19740-S2 and 19797-S2; Part Number: 310-0006 (Lat Assembly, Rad Stem Implant, Collar 6mm, size 3) Lot #'s 23199101, 23425001, 19059-S2, 19680-S2, 19741-S2, 19798-S2 and 19845-S2; Part Number: 310-0007 (Lat Assembly, Rad Stem Implant, Collar 6mm, size 4) Lot #"s 22890901, 2345101, 19060-S2, 19125-S2 and 19457-S2; Part Number: 310-0008 (Lateral Assembly, Radial Head Implant, Size 2) Lot #'s 22288701, 22629401, 22891001, 23199201, 23425201 and 23917401; Part Number:(Lateral Assembly, Radial Head Implant, Size 3) Lot #'s 15886, 22288801, 22629501, 22891101, 23199301, 23425301, 23917501 and 23917502; Part Number: 310-0010 (Lat Assembly, Rad Stem Head Implant, Size 4) Lot #'s 15887, 23199401, 23425401, 23917601 and 19458-S2; Part Number: 310-0011 (Lateral Assembly, Radial Head Impl Assembly, Size 2) Lot #'s 16708, 16709, 22629601, 22891201 and 23425501; Part Number: 310-0012 (Lateral Assembly, Radial Head Impl Assembly, Size 3) Lot #'s 16710, 16711, 22629701, 22891301 and 23425601; Part Number: 310-0013 (Lateral Assembly, Radial Head Impl Assembly, Size 4) Lot #'s 16712 and 23425701; Part Number: 310-2010 (rHead Stem Implant Plasma Coated, Size1) Lot #'s 22552401, 19693-S2, 19742-S2 and 19799-S2; Part Number: 310-2011 (rHead Stem Implant Plasma Coated, Size 2) Lot #'s 22552501, 19694-S2 and 19743-S2; Part Number: 310-2012 (rHead Stem Implant Plasma Coated, Size 3) Lot #'s 22552601, 19070-S2, 19695-S2, 19744-S2 and 19800-S2; Part Number: 310-2013 (rHead Stem Implant Plasma Coated, Size 4) Lot #'s 22552701, 19696-S2 and 19745-S2; Part Number: 310-2014 (rHead Stem Implant 6mm Collar, Size 1) Lot #'s 22552801, 22552802, 19697-S2, 19698-S2 and 19746-S2; Part Number: 310-2015 (rHead Stem Implant 6mm Collar, Size 2) Lot #'s 22552901, 19699-S2, 19747-S2 and 19915-S2; Part Number: 310-2016 (rHead Stem Implant 6mm Collar, Size 3) Lot #'s 22553001, 19700-S2 and 19748-S2; Part Number: 310-2017 (rHead Stem Implant 6mm Collar, Size 4) Lot #22553101; Part Number: 310-2018 (rHead Recon Stem Implant Plasma Coated, Size 1) Lot #'s 22553201, 22889801, 19701-S2, 19749-S2 and 19801-S2; Part Number: 310-2019 (rHead Recon Stem Implant Plasma Coated, Size 2) Lot #'s 22553301, 22889901, 19702-S2, 19750-S2, 19916-S2 and 19975-S2; Part Number: 310-2020 (rHead Recon Stem Implant Plasma Coated, Size 3) Lot #'s 22553401, 22890001, 19703-S2, 19751-S2, 19802-S2 and 19976-S2; Part Number: 310-2021 (rhead Recon Stem Implant Plasma Coated, Size 4) Lot #'s 22553501, 22890101, 197076-S2, 19704-S2, 19752-S2, 19917-S2 and 19977-S2; Part Number: 310-2022 (rHead Recon Stem Implant 6mm Collar, Size 1) Lot #'s 22553601, 22890201, 197501-S2 and 19753-S2; Part Number: 310-2023 (rHead Recon Stem Implant 6mm Collar, Size 2) Lot #'s 22553701, 22890301, 19706-S2 and 19754-S2; Part Number: 310-2024 (rHead Recon Stem Implant 6mm Collar, Size 3) Lot #'s 22553801, 22890401, 19075-S2, 19707-S2 and 19755-S2; Part Number: 310-2025 (rHead Recon Stem Implant 6mm Collar, Size 4) Lot #'s22553901, 22890501 and 19071-S2;  Part Number: 410-0100 (rHead Standard Extended Stem 6mm Collar, Size 1)  Lot #'s 18527, 18724, 19681-S2 and 19715-S2; Part Number: 410-0101 (rHead Standard Extended Stem 6mm Collar, Size 2) Lot #'s 18528, 18725, 19682-S2 and 19756-S2; Part Number: 410-0102 (rHead Standard Extended Stem 6mm Collar, Size 3) Lot #'s 18529, 18726, 19663-S2 and 19757-S2; Part Number: 410-0103 (rHead Standard Extended Stem 6mm Collar, Size 4) Lot #'s 18530, 18727, 19683-S2 and 19758-S2;  Part Number: RCN-S160 (rHead Recon Stem Implant Non-Coated, Size 1) Lot #'s 18370, 18371, 18388, 18569, 18698, 18718 22614101, 22849001, 23068201, 19102-S2, 19103-S2, 19104-S2, 19717-S2, 19804-S2 and 19920-S2; Part Number: RCN-S260 (rHead Recon Stem Implant Non-Coated, Size 2) Lot #'s 18372, 18373, 18699, 18719, 18840, 22614301, 22849201, 23068401, 19105-S2, 19106-S2, 19107-S2, 19773-S2, 19851-S2 and 19984-S2; Part Number: RCN-S360 (rHead Recon Stem Implant Non-Coated, Size 3) Lot #'s 18570, 18571, 18700, 18701, 18702, 18755, 18841, 1884222298301, 22298302, 22849401, 23479801, 19108-S2, 19723-S2, 19807-S2, 19854-S2, 19986-S2 and R23479801; Part Number: RCN-S460 (rHead Recon Stem Implant Non-Coated, Size 4) Lot #'s18572, 18703, 18704, 18756, 18843, 18844, 22298401, 22614601, 22849601, 19109-S2 and 19110-S2; Part Number: RHA-S160 (rhead Radial Implant 6mm Collar, Size 1) Lot #''s 18375, 18386, 18501, 18707, 18708, 18848, 18849, 18850, 18851, 18852, 22613101, 22850101, 23069101, 19113-S2, 19114-S2, 19115-S2, 19779-S2 and 19809-S2; Part Number: RHA-S260 (rhead Radial Implant 6mm Collar, Size 2) Lot #'s 18376, 18384, 18385, 18502, 18575, 18709, 18720, 18853, 18854, 22613301, 22850301, 23069301, 19116-S2, 19117-S2 and 19782-S2; Part Number: RHA-S360 (rhead Radial Implant 6mm Collar, Size 3) Lot #'s 18377,18383, 18503, 18710, 18711, 18856, 18857, 18858, 18859, 22613501, 22850501, 19118-S2, 19119-S2, 19730-S2 and 19813-S2; Part Number: RHA-S460 (rhead Radial Implant 6mm Collar, Size 4) Lot #'s 18379, 18382, 18504, 18576, 18712, 18713, 18860, 18861, 18862, 22613701, 22850701, 19120-S2 and 19121-S2

Why it was recalled

Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.

Root cause (FDA determination)

Packaging change control

Action the firm took

Branches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Dr, Mahwah, New Jersey 07430-2006

Distribution

Distribution pattern
Worldwide Distribution.

Timeline

Recall initiated
2015-06-24
Posted by FDA
2015-08-20
Terminated
2016-04-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #138723. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.