Recalls / —
—#138751
Product
Radio-Capitellum, Capitellum, Left, Large, Do Not Reuse, Rx Only. Indicated for use in the elbow for reduction or relief of pain and/or improved elbow function in skeletally mature patients.
- FDA product code
- JDB — Prosthesis, Elbow, Semi-Constrained, Cemented
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K070236
- Affected lot / code info
- Part Number: 410-0000 (Radio Capitellum Large, Left) - Lot #'s 18832, 22193201, 22193202, 22193203, 22258301,22957101, 23853901, 19069-S2, 19077-S2 and 19155S2; Part Number: 410-0001 (Radio Capitellum Small, Left) - Lot #'s18833, 18834, 22193302, 22193303, 22766801, 2295601 and 23853101; Part Number: 410-0002 (Radio Capitellum Large, Right) - Lot #'s18835, 18836, 22193401, 22193402, 22193403, 22766901, 22956301, 23853201 and 19074-S2; Part Number: 410-0003 (Radio Capitellum Small, Right) - Lot #'s 18565, 22193501, 22193502, 22193503, 22767001, 22956401 and 23853301
Why it was recalled
Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.
Root cause (FDA determination)
Packaging change control
Action the firm took
Branches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.
Recalling firm
- Firm
- Stryker Howmedica Osteonics Corp.
- Address
- 325 Corporate Dr, Mahwah, New Jersey 07430-2006
Distribution
- Distribution pattern
- Worldwide Distribution.
Timeline
- Recall initiated
- 2015-06-24
- Posted by FDA
- 2015-08-20
- Terminated
- 2016-04-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #138751. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.