Product
Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.
- FDA product code
- JFY — Enzymatic Method, Creatinine
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K061238, K925668
- Affected lot / code info
- Dimension CREA Lot Numbers: DB5195 FA5211 EA5233 DA5258 FA5275 BA5293 EA5315 EA5316 GA5342 FA5351 FA5352 FA6006 BA6026 BA6027 EB6047 GB6076 GA6097 BA6111 BA6112 FB6139 with expiration dates: 2015-07-14 2015-07-30 2015-08-21 2015-09-15 2015-10-02 2015-10-20 2015-11-11 2015-11-12 2015-12-08 2015-12-17 2015-12-18 2016-01-06 2016-01-26 2016-01-27 2016-02-16 2016-03-16 2016-04-06 2016-04-20 2016-04-21 2016-05-20 Dimension Vista CREA Lot Numbers: 14177BB 14212BD 14234BH 14261BC 14280BA 14301BB 14321AC 14346AC 15012AA 15042BA 15054AB 15076BA 15091BA 15125BB 15125BA 15141AA with expiration dates: 2015-06-26 2015-07-31 2015-08-22 2015-09-18 2015-10-07 2015-10-28 2015-11-17 2015-12-12 2016-01-12 2016-02-11 2016-02-23 2016-03-16 2016-03-31 2016-05-04 2016-05-04 2016-05-22
Why it was recalled
Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.
Root cause (FDA determination)
Device Design
Action the firm took
Siemens sent an Urgent Medical Device Correction letter, dated July 2015, was provided to all Dimension CREA and Dimension Vista CREA customers who received the products to notify them of the issue, the potential risk to health, and actions to be taken by customers. If you have any questions please contact your Siemens Customer Care Center Technical Solutions at (800) 441-9250, or your local Siemens Technical Support Representative.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- Worldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Great Britain, Greece, Guatemala, Hungary, Iraq, Ireland, Honduras, Hong Kong, Iceland, Indonesia, India, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Libya, Macedonia, Malaysia, Mali, Mexico, Myanmar, Morocco, Netherlands, Nepal, Nicaragua, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Uruguay, Venezuela, Vietnam and Yemen.
Timeline
- Recall initiated
- 2015-07-15
- Posted by FDA
- 2015-09-05
- Terminated
- 2017-03-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #138835. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.