Recalls / —
—#139007
Product
SHOULDER PACK 6/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .
- FDA product code
- OJH — Orthopedic Tray
- Device class
- Class 1
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- Code:9002616 Lots: 141015729 exp. 10/31/15 141216309 exp. 12/31/15 150116557 exp. 2/28/16 150317941 exp. 3/31/16 150317942 exp. 3/31/16 150317943 exp. 4/30/16 150418380 exp. 5/31/16 150317945 exp. 5/31/16 150317944 exp. 5/31/16 150418381 exp. 5/31/16 150418379 exp. 5/31/16 150317946 exp. 6/30/16
Why it was recalled
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Root cause (FDA determination)
Packaging
Action the firm took
Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Recalling firm
- Firm
- Customed, Inc Calle Igualdad Final # 7
- Address
- Fajardo, Puerto Rico 00738
Distribution
- Distribution pattern
- Distributed Only in Puerto Rico.
Timeline
- Recall initiated
- 2015-07-10
- Posted by FDA
- 2015-09-03
- Terminated
- 2017-04-12
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #139007. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.