Product

C-SECTION PK 2/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate

FDA product code
OHMCesarean Section Tray
Device class
Class 2
Medical specialty
Obstetrics/Gynecology
Affected lot / code info
Code: 900998 Lots: 141015749 exp. 10/31/15 141015749 exp. 10/31/15 140915186 exp. 11/30/15 150116835 exp. 3/31/16 150317856 exp. 3/31/16 150317857 exp. 3/31/16 150518853 exp. 5/31/16

Why it was recalled

Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Root cause (FDA determination)

Packaging

Action the firm took

Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.

Recalling firm

Firm
Customed, Inc Calle Igualdad Final # 7
Address
Fajardo, Puerto Rico 00738

Distribution

Distribution pattern
Distributed Only in Puerto Rico.

Timeline

Recall initiated
2015-07-10
Posted by FDA
2015-09-03
Terminated
2017-04-12
Status

Source: openFDA Device Recall endpoint. Recall record ID #139164. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.