Recalls / —
—#139865
Product
DeRoyal Sterile Custom Surgical Kits containing 3M DuraPrep (TM) Surgical Solution, Catalog No. 8630, Lot Nos: 2016-09EE: TOTAL KNEE A & B, REF 89-4561.11; TOTAL HIP A & B, REF 89-4562.12; PODIATRY TRAY, REF 89-3101.04; KNEE SCOPE PACK, REF 89-7744.03; KNEE PACK, REF 89-8194.02; BEACH CHAIR PACK, REF 89-8439.01; TOTAL JOINT, REF 50-12160.17; BACK TRAY, REF 50-12602.09; PACEMAKER PACK, REF 89-8381.01;
- FDA product code
- LRO — General Surgery Tray
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- REF 89-4561.11, Lot numbers: 38198163, 38276045, 38276045; REF 89-4562.12, Lot Number 38345071; REF 89-3101.04, Lot Number 38276230; REF 89-7744.03, Lot Number 38229177; REF 89-8194.02, Lot Number 38245020; REF 89-8439.01, Lot Number 38317548; REF 50-12160.17, Lot Number 38247391; REF 50-12602.09, Lot Number 38345266; REF 89-8381.01, Lot Number 38085071
Why it was recalled
Custom Sterile Surgical kits contain 3M DuraPrep (TM) Surgical Solution, Catalog Number 8630, Lot Number 2016-09EE which was recalled due to unintended exposure to the sterilant during package sterilization.
Root cause (FDA determination)
Other
Action the firm took
DeRoyal initiated their recall on 08/31/2015 to the end-user level by sending overnight letters to its direct accounts and hospitals. Distributors were asked to put the product in quarantine, place the provided labels on the products and complete and return the spreadsheet of how many they labeled. They were asked to return the spreadsheet even if they did not have inventory. Hospitals were provided with labels to affix to the kits with direction to discard the recalled component and notice of destruction forms to be completed and returned.
Recalling firm
- Firm
- DeRoyal Industries Inc
- Address
- 200 Debusk Ln, Powell, Tennessee 37849-4703
Distribution
- Distribution pattern
- SC, FL, OH, NJ, OK, MN, AR
Timeline
- Recall initiated
- 2015-08-31
- Posted by FDA
- 2015-11-06
- Terminated
- 2016-04-12
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #139865. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.