Product
Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
- FDA product code
- GXN — Plate, Cranioplasty, Preformed, Non-Alterable
- Device class
- Class 2
- Medical specialty
- Neurology
- 510(k) numbers
- K992000
- Affected lot / code info
- Part Number 329.315 460.001 460.001.01S 460.002 460.002.01S 460.003 460.003.01S 460.008 460.008.01S 460.009 460.009.01S 460.010 460.010.01S 460.100 460.100.01S 460.107 460.107.01S with Lot Numbers 4146716, 4154144, 4206682, 4206683, 4213557, 4214824, 4216603, 4216604, 4240689, 4240690, 4241741, 4248462, 4249395, 4249396, 4264386, 4278302, 4278304, 4283035, 4283036, 4283037, 4283038, 4283039, 4283040, 4283041, 4285956, 4289103, 4295084, 4295085, 4298044, 4303093, 4326975, 4326979, 4326981, 4329355, 4331263, 4331265, 4331266, 4344659, 4344678, 4378973, 4482201, 4485317, 4488093, 4488094, 4490608, 4490609, 4496716, 4496915, 4496916, 4501971, 4504043, 4508068, 4513458, 4513459, 4513460, 4513461, 4516371, 4519406, 4519407, 4521408, 4525905, 4550062, 4550063, 4550064, 4550065, 4551164, 4551165, 4551166, 4561400, 4561401, 4572176, 4574481, 4580884, 4586704, 4605094, 4661414, 4669482, 4669483, 4675725, 4695343, 4705970, 4723148, 4753792, 4753794, 4777999, 4778000, 4906228, 4906229, 4914498, 4917681, 4934562, 4944809, 4956361, 4956363, 4983255, 4983256, 4983257, 5007806, 5041856, 5041857, 5041858, 5046857, 5046859, 5098159, 5113496, 5113497, 5134868, 5134869, 5143883, 5204105, 5230669, 5234395, 5234396, 5245233, 5245237, 5251710, 5292433, 5308690, 5308733, 5341509, 5370371, 5370375, 5438364, 5452452, 5465222, 5655435, 5667092, 5697537, 5726966, 5726967, 5726968, 5726969, 5726970, 5726971, 5801281, 5801284, 5801285, 5801287, 6036379, 6036381, 6168066, 6188888, 6188890, 6268389, 6268390, 6272370, 6322541, 6364587, 6553805, 6861415, 6983782, 7032568, 7505227, 7640627
Why it was recalled
The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
Root cause (FDA determination)
Labeling False and Misleading
Action the firm took
DePuy Synthes sent an URGENT:FIELD SAFETY NOTIFICATION / MEDICAL DEVICE LABELING CORRECTION letter, dated August 31, 2015 to affected customers. The letter identified affected product, problem and actions to be taken. Customers were instructed to: "Update and review the package insert (GP1345-D). " Discard outdated package insert noted in the table above. " Forward this Field Safety Notification to anyone in your facility that needs to be informed, especially personnel that conduct MR testing. " If the package insert has been forwarded to another facility, contact that facility. " Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. Send a copy of the completed Verification Section by, Fax: 877-884-7821 or Scan/email: Synthes3911@stericycle.com " If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification. " Maintain a copy of this notice. For questions call 610-719-5450.
Recalling firm
- Firm
- Synthes (USA) Products LLC
- Address
- 1301 Goshen Pkwy, West Chester, Pennsylvania 19380-5986
Distribution
- Distribution pattern
- US Nationwide and internationally, including Canada.
Timeline
- Recall initiated
- 2015-08-31
- Posted by FDA
- 2015-09-23
- Terminated
- 2016-08-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #139935. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.